Easy Chem
Clinical Chemistry Analyser
₹130000.00₹100000.00
Technical Specifications
PHOTOMETRIC MODULE
Filter Wavelengths : 340, 405, 505, 546, 578, 630 and 670nm (7+ 1 open position for optional filter)
IAD process ensures long filter life
Wavelength Accuracy : + 2mm
Silicon Photo Diode detector
Absorbance Range : 0.0-3.0A
Resolution : 0.0001 Absorbance Units
Photometric Drift : < 5 mAbs / 30 min
CV : < 1%
ANALYSIS METHOD
Endpoint, Fixed time and Kinetic
Sample blanking and Reagent blanking
Direct Absorbance measurement mode
Coagulation Mode (PT / APTT / TT)
TEST PROFILES
User defined 10 Test Profies
10 tests can be added in each profile
TEMPARATURE OF FLOW CELL / CUVETTE
Temparature Stability : + 0.2ºC
25ºC, 30ºC, 37ºC
TEST CHANNELS
Direct access keys for dedicated assays
48 dedicated assays
Open channel capability
200 programmable channels
DISPLAY
High Contrast Display
5.7 Inch LCD display screen
DATA MANAGEMENT
Stores 30 days QC for each Assay
Storage of up to 10,000 test results
Reagent Blank storage
KEYBOARD
Visualindication on keypad during aspiration
Tactile touch QWERTY keypad
PRINTER
Internal 57 mm thermal printer with full graphics
printing, maintenance free
MEASURING MODULE
High quality Quartz flow cell for precise measurements
Dual mode (Flow cell & Cuvette)
30ul Flow cell volume
Peltier-controlled reaction temparature
External cuvettes (10mm path length)
UNITS
15 pre-defined measurement units
REACTION VOLUME
Maximum aspiration volume - 5ml
Minimum aspiration volume - 300 ul
DIMENSIONS
Width : 324mm
Length : 422mm
Height : 176mm
Weight : 6.5kg
REPORTS
Calibration Reports QC results
Search by date, sample ID or Test Name
INTERNAL INCUBATOR
Temparature : 37 Degree
Position : 7
QC AND CALIBRATION
Daily and monthly QC with data archiving
INTERFACE
PC Connectivity (optional)
RS 232 Por
USB Port
POWER
Input Voltage Range : 88-264 VAC
Frequency Range : 47-63 Hz
Power consumption : 75 W max
Brand
Spectrum Biomedical And Engineering Services
Contact
Newsletter
biomedical.spectrum@email.com
+91-8116 261 073
© 2024. All rights reserved.
Address
REGD. OFFICE: LODNA, ONDA, BANKURA 722144
STOCK: DEBANGAN BUILDING, BHOWMIK GATE, NATUNCHATI, BANKURA 722101